> Products > Constant temperature and humidity room
|
|
 |
|
|
Molecular Phenom Analysis Room(incubation type - 1) |
|
Specification - Molecular phenome analysis room
|
|
Product Name
|
Specifications
|
HNS-Bio6CRH
|
A: Specifications
|
1. In Size Dimension : 5,500 X 8,000X 2,700 (m/m)
|
B: Temperature
|
1. Temperature Range: 10 ~ 30¡É (18¡É± 0.2¡É)
|
2. Humidity Range:
40% ~ 99% (50%± 2%)
|
3. Temperature Control : MICROPROCESSOR BASED P.I.D. CONTROL
|
4. Use cleanliness
: Class 100,000
|
5. Air circulation device
: UNIFORM DOWNWARD AIR FLOW.
|
6. Alarm Function: Deviation in the range of temperature and
humidity settings registered during Your mobile phone (n = 4) of the alarm transmission
|
7. Cooling System : HERMETICALLY SEALED AIR COOLED
CONDENSING UNIT. 5 HP 1SET
|
8. Temperature Control Type: Hot Gas By Pass Type
|
9. LAMP : 32W Three-wavelength LAMP 18SET
|
C: HEPA Filter Box
|
1. HEPA Filter (m/m) : 610*610*150, 0.3§ 99.97%
|
2. Quantity
: 8 SET
|
D: Insulation
|
1. INTERIOR : Polyester Painted Steel Sheet
|
2. Insulation :Foam Insulation 75t
|
3. EXTERIOR : Polyester Painted Steel Sheet
|
E: Power
|
AC 380V 3 PH 4W KW (MAX)
|
|
|
|
Molecular Phenom Analysis Room(incubation type - 2) |
|
Specification - Automated Molecular Phenome Room
|
|
Product Name
|
Specifications
|
HNS-Bio7CRH
|
A: Specifications
|
1. In Size Dimension : 7,000 X 8,000X 2,700 (m/m)
|
B: Temperature
|
1. Temperature Range: 15 ~ 30¡É (22¡É± 0.2¡É)
|
2. Temperature Control : MICROPROCESSOR BASED P.I.D. CONTROL
|
3. Use cleanliness
: Class 100,000
|
4. Air circulation device
: UNIFORM DOWNWARD AIR FLOW.
|
5. Alarm Function: Deviation in the range of temperature and
humidity settings registered during Your mobile phone (n = 4) of the alarm transmission
|
6. Cooling System : HERMETICALLY SEALED AIR COOLED
CONDENSING UNIT. 5 HP 1SET
|
7. Temperature Control Type: Hot Gas By Pass Type
|
8. LAMP : 32W Three-wavelength LAMP 18SET
|
C: HEPA Filter Box
|
1. HEPA Filter (m/m) : 610*610*150, 0.3§ 99.97%
|
2. Quantity
: 6 SET
|
D: Insulation
|
1. INTERIOR : Polyester Painted Steel Sheet
|
2. Insulation :Foam Insulation 75t
|
3. EXTERIOR : Polyester Painted Steel Sheet
|
E: Power
|
AC 380V 3 PH 4W KW (MAX)
|
|
|
|
Molecular Phenom Analysis Room(incubation type - 3) |
|
Specification - Plant Sample
Preparation
Room
|
|
Product Name
|
Specifications
|
HNS-Bio8RH
|
A: Specifications
|
1. In Size Dimension : 3,400 X 7,000 X 2,700 (m/m)
|
B: Temperature
|
1. Temperature Range: -10 ~ 20¡É (4¡É± 0.2¡É)
|
2. Humidity Range: 30% ~ 99%
(50%± 2%)
|
3. Temperature Control : MICROPROCESSOR BASED P.I.D. CONTROL
|
4. Alarm Function: Deviation in the range of temperature and
humidity settings registered during Your mobile phone (n = 4) of the alarm transmission
|
5. Cooling System : HERMETICALLY SEALED AIR COOLED
CONDENSING UNIT. 3 HP 2SET(Double cooling)
|
6. Temperature Control Type: Hot Gas By Pass Type
|
7. LAMP : 18W Three-wavelength LAMP 6 SET
|
D: Insulation
|
1. INTERIOR : Polyester Painted Steel Sheet
|
2. Insulation :Foam Insulation 75t
|
3. EXTERIOR : Polyester Painted Steel Sheet
|
E: Power
|
AC 380V 3 PH 4W KW (MAX)
|
|
|
|
|
|
Model
|
HNS - Bio9CRH
|
In Side Dimension
|
7000 * 8000 * 2700 m/m
|
Temperature Range
|
0¡É ~ 40¡É
|
Temperature difference
|
25¡É ± 0.3¡É Within
|
Temperature error
|
0.1¡É Within
|
Humidity
|
0 ~ 99%
|
Humidity error
|
25¡É ± 2.5%
|
Dehumidifier
|
Cooling System
|
Co2 Device
|
Range 0 ~ 5000 ppm
|
Heating device
|
Hot Gas By Pass System
|
Illumination Devices
|
8 STEP 0 ~ 25,000 LUX
|
Pressure
|
Auto Damper
|
The filter unit
|
Hepa Filter Unit(Class100,000)
|
Amount of liquid
|
m/h
|
Control
|
24-step Digital PID Controller(temperature and humidity, illuminance, Co2, exhaust, ventilation)
|
Product Material
|
Interior : Polyester Painted Steel Sheet
External : SUS or C.G.I Steel plate, flame-proof, acid-treated
|
Power Consumption
|
W
|
Power
|
3¡Ë4w 380V 60(50)Hz
|
Protection
|
Undercooling / overheating protection devices, water level control, over-current protection device, phase, reverse phase, under-voltage protection device
|
|
|
- All products can be made to order.
|
|
|
Chamber Room (Chamber Type) |
|
|
MODEL
|
HCR-2500
|
HCR-3500
|
HCR-4000
|
In Size Dimension
( W X D X H )mm
|
2500*2500*2500
|
3500*3000*2500
|
1000*800*1000
|
Out Size Dimension
( W X D X H )mm
|
3450*2750*2825
|
4450*3250*2825
|
4950*4250*2825
|
Temperature Range
|
-40 ~ 80¡É
|
Humidity Range
|
30 ~ 99%
(¡ØHourly temperature prior consultation period is required for humidity control)
|
Temperature Constancy
|
±1.0¡É
|
Humidity Constancy
|
±3.0¡É
|
Material
|
In
|
SUS304 1.5T Polishing
|
Out
|
SS41 Coating
|
Insulation
|
Poly-Urethane
|
Heating Time
|
RT ~ 70¡É (30min)
|
Cooling Time
|
RT ~ -30¡É (60min)
|
Heating System
|
Coil Heater(Temperature) & Pipe Heater(Humidity)
|
Refrigeration System
|
The outdoor-type hard, binary refrigeration system
(air-cooled and water-cooled optional)
|
Safety Devices
|
Thermal cut protector, refrigerator overload circuit breakers, motor overload circuit breaker, single main tank and humidifier tank protection devices, external alarm device
|
View Window
|
3Pair Heated Glass No Dew Condensation
|
Sensor
|
Electronic humidity sensor
|
Controller
|
Program Controller (Touch Screen Type)Interface : RS232 or RS485
|
Recorder
(Option)
|
6Chanel (Paper) or 3,6,12Chanel Digital Type(Paperless)
|
Power Source
|
AC 220V / 380V 3 ¥Õ 4w
|
|
|
|
|
|
 |
|